PCR Laboratory

PCR laboratory

1.Clean bench; 2. Medicine refrigerator; 3. UV Disinfection  trolley; 4. Low temperature refrigerator; 5. Metal  bath; 6. Centrifuge; 7. Water bath; 8. Nucleic extractor; 9. Pipette; 10. biosafety cabinet; 11. PCR machine; 12. autoclave; 13.Vortex mixer

area product function quantity brand Model
Reagent preparation area Clean bench Configure reagents 1 OLABO BBS-SDC
Centrifuge Centrifuge samples 1 OLABO Mini-12
Vortex mixer Mix the sample 1 OLABO 88882010
Metal bath Reagent dissolving and heating 1 OLABO 88870005
Pipette Pipetting 4 OLABO 0.5-10µl
10-100μl
20-200μl
100-1000μl
Pipette holder Place the pipette 1 OLABO Linear
Low temperature refrigerator Store reagents 1 OLABO BDF-25V270
Medicine refrigerator Store reagents 1 OLABO BYC-310
UV disinfection Trolley Space disinfection 1 OLABO MF-Ⅱ-ZW30S19W
Biosafety transport box Sample transportation 1 OLABO QBLL0812
Specimen preparation area Biological safety cabinet Sample processing 1 OLABO BSC-1500IIB2-X
Water bath Sample inactivation 1 OLABO HH-W600
Vortex mixer Mix the sample 1 OLABO 88882010
Centrifuge Sample centrifugation 1 OLABO TG-16W
TGL-16M
OLABO Mini-12
Pipette Pipetting 4 OLABO 0.5-10µl
10-100μl
20-200μl
100-1000μl
Pipette holder Place the pipette 1 OLABO Linear
Nucleic Acid Extractor Extract nucleic acid 1 OLABO BNP96
Low temperature refrigerator Sample storage 1 OLABO BDF-86V348
Medicine refrigerator Reagent storage 1 OLABO BYC-310
UV disinfection Trolley Space disinfection 1 OLABO MF-Ⅱ-ZW30S19W
amplification analysis area PCR machine Sample amplification test 1 OLABO MA-6000
Low temperature refrigerator Store samples 1 OLABO BDF-25V270
UV disinfection Trolley Space sterilization 1 OLABO MF-Ⅱ-ZW30S19W
Disinfection area Autoclave Disinfection of surgical instruments 1 OLABO BKQ-B75II
Experimental consumables Tip Use with pipette According to actual needs OLABO TF-100-R-S
TF-1000-R-S
TF-300-R-S
TF-200-R-S
PCR tube Use with a fluorescent quantitative PCR instrument According to actual needs OLABO PCR-0208-C
PCR-2CP-RT-C
Centrifuge tube Store reagent samples or use with centrifuge According to actual needs OLABO MCT-150-C
Sampling tube Collect samples 1 OLABO
Nucleic Acid Extraction Kit Use with nucleic acid extractor According to actual needs OLABO
Protective consumables Medical masks Protective Equipment According to actual needs OLABO
protective suit Protective Equipment According to actual needs OLABO 
Disposable isolation gown Protective Equipment According to actual needs OLABO
Gloves Protective Equipment According to actual needs OLABO
Alcohol Disinfection supplies According to actual needs OLABO 500ml
Hand sanitizer Disinfection supplies According to actual needs OLABO 500ml

PCR laboratory

1. P2 sterile room (excluding buffer room) is equipped with a set of clean air-conditioning system.

2. The storage room for one-time change of shoes and rain gear, the secondary dressing room, the washing room and the buffer room are each equipped with a split purification air-conditioning system and exhaust fan.

3. The outdoor fresh air enters the computer room through the primary filter, and then is sent to the high efficiency filter (supply vent) on the top of the clean room through the air conditioner, medium-efficiency filter, and pressurized fan. Each vent is equipped with air volume adjustment. Valve for air volume distribution and pressure difference adjustment.

4. The air outlet at the end of the air exhaust system is 0.3 meters above the ground, and there is an air volume regulating valve for air volume adjustment. The exhaust main pipe must be installed with a non-return room to prevent the exhaust air from flowing backward.
5. The temperature probe is set in the main return air duct.