CPR machine races against life

There is only one life. But before the crisis comes, Death did not give you time to prepare and regret. All you can do is race against time and life. Correct and timely CPR may save the lives of your loved ones.

Cardiopulmonary resuscitation (CPR), also known as basic life support, is one of the first aid measures that must be taken within 4 to 6 minutes for cardiac arrest due to various reasons. The purpose is to rescue brain cells from necrosis in hypoxic state as soon as possible (more than four minutes to cause brain damage, and more than ten minutes to cause irreversible damage to the brain), so the sooner the rescue time, the better. Cardiopulmonary resuscitation (CPR) is suitable for unconsciousness and respiratory and cardiac arrest caused by heart attack, drowning, suffocation or other accident.

Theoretically, for patients with sudden cardiac death, there is a positive relationship of approximately 10% per minute, and the success rate of CPR within 1 minute of cardiac arrest is greater than 90%. If the cardiac arrest is more than 10 minutes, CPR is performed again, and the corresponding success rate is almost 0.Relationship between CPR start time and success.

未标题-3

The CPR machine only takes ten seconds.

Overview

Cardiopulmonary resuscitation machine is one of the most widely used emergency equipment in emergency and ICU. It can provide advanced, reliable, continuous and ideal cardiopulmonary resuscitation for patients with cardiac and respiratory arrest at the first time, and can be implemented safely, quickly and effectively rescue.

 心肺复苏机

Effect

Continuously perform standardized CPR techniques.

Relieve rescuer fatigue.

When ambulance personnel are limited, trainees can be freed up to implement advanced life support measures.

During the transfer, it can ensure that the patient receives continuous and uninterrupted correct compression.

The risk of cross-infection between doctors and patients due to mouth-to-mouth blowing is excluded.

心肺复苏机彩页

 parameters:

1、Pressing depth:30 ㎜~65 ㎜ adjustable,one type:5 ㎜,error:±2 ㎜;
2、Pressing release rate:1:1 ;
3、pressure relief valve:0.65Mpa;error:±0.12Mpa;
4、Pressure of gas road is no less than 1Mpa;
5、Ventilation quanity:500ML no adjust,error:±10%;
                                        working pressure:≤816.4N ;
6、input power(DC):≤17W; AC:≤54W;
7、Press the ventilation ratio:30:2 continuously adjustable;,error:±1 time;
8、Pressing frequency: 100times/ min,110times/ min,120times/ min error:±1 time;
9、The fastness of fastening band fastness is not less than 100N;
10、The thickness range of patients was 155 to 305MM.
11、The pressure is not enough to alarm, the pressure depth is reduced automatically, and the manual recovery after air pressure recovery must be deep.
12、Electricity, according to the low voltage alarm, start the suspended function;
13、Default mode:Press the ventilation ratio: 30: 2,press depth 30mm,press frequence 100 pcs/min;Ventilation quanity 500ML
14、Safety requirements refer to medical electrical appliances safety standard GB 9706.1-2007, electromagnetic compatibility standard YY05052012;
15、Earthquake Resistance
16、drop resistance
17、The enclosure protection level is IP55 (dustproof and water spray);
18、noise≤70dB;
19、Working environment :
a)Temperature:-20℃~55℃;
b)Relative temperature:10%~90%;
c)Atmospheric pressure:500hPa~1060hPa.