Personlized Products Laminar Chamber - Mini PCR Work Station – OLABO
Personlized Products Laminar Chamber - Mini PCR Work Station – OLABO Detail:
Advantage
1.The multi-function console is divided into three areas, and the areas do not interfere with each other; at the same time, the three areas are closely designed to reduce the sample transfer time between areas and improve the efficiency of rapid nucleic acid detection.
2.The reagent preparation area is a positive pressure environment, which can protect the reagent preparation from external pollution and protect the reagent.
3.The specimen preparation area is a negative pressure environment, which can not only protect personnel from harmful aerosol infection, but also protect samples from external pollution, and also protect the external environment from pollution.
4.The amplification analysis area is a negative pressure environment, which can not only protect personnel from harmful aerosol infection, but also protect the external environment from pollution
PCR laboratory
1.Clean bench; 2. Medicine refrigerator; 3. UV Disinfection trolley; 4. Low temperature refrigerator; 5. Metal bath; 6. Centrifuge; 7. Water bath; 8. Nucleic extractor; 9. Pipette; 10. biosafety cabinet; 11. PCR machine; 12. autoclave; 13.Vortex mixer
PCR laboratory
1. P2 sterile room (excluding buffer room) is equipped with a set of clean air-conditioning system.
2. The storage room for one-time change of shoes and rain gear, the secondary dressing room, the washing room and the buffer room are each equipped with a split purification air-conditioning system and exhaust fan.
3. The outdoor fresh air enters the computer room through the primary filter, and then is sent to the high efficiency filter (supply vent) on the top of the clean room through the air conditioner, medium-efficiency filter, and pressurized fan. Each vent is equipped with air volume adjustment. Valve for air volume distribution and pressure difference adjustment.
4. The air outlet at the end of the air exhaust system is 0.3 meters above the ground, and there is an air volume regulating valve for air volume adjustment. The exhaust main pipe must be installed with a non-return room to prevent the exhaust air from flowing backward.
5. The temperature probe is set in the main return air duct.
Product detail pictures:






Related Product Guide:
Our crew through skilled training. Skilled skilled knowledge, strong sense of company, to meet the company wants of customers for Personlized Products Laminar Chamber - Mini PCR Work Station – OLABO, The product will supply to all over the world, such as: Los Angeles, Iraq, Bhutan, Our solutions have national accreditation standards for experienced, premium quality items, affordable value, was welcomed by people around the globe. Our goods will continue to increase in the order and look forward to cooperation with you, Really should any of those products be of interest to you, please letus know. We are going to be pleased to give you a quotation up on receipt of one's detailed specifications.