Parameter
Model | BK-400 | |
Overall | Throughput | 400 Test/hour |
Performance | Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry |
Certification | CE, ISO9001, ISO14001, ISO13485 | |
Sample Positions | 60 or 90 sample positions, built-in barcode system available | |
Sample & Reagent System | Reagent Positions | 45 refrigerated reagent positions |
Sample Volume | 2~70μl | |
Reagent Volume | 20~350μl | |
Probe | Teflon coating, with anti-collision function, liquid leveldetection function | |
Probe Washing | Automatic washing interior and exterior | |
Reagent Cooling | Refrigerated tray with independent switch | |
Temperature Control | 37±0.1℃, real-time monitoring | |
Reaction System | Cuvettes | 120reusable cuvettes, optical length 6mm |
Mixer Probe | Independent stirring | |
Washing | Automatic cuvettes washing | |
STAT function | YES | |
Optical System | Light Source | 12V/30W halogen-tungsten lamp |
Spectrophotometry | Post-spectral spectrophotometry | |
Wavelength | 340, 405, 450,510, 546, 578, 630,700nm | |
Absorbance | 0~3.0Abs | |
Calibration & QC | Calibration | Linear:K factor, 1-point, 2-point and multipoint linear |
Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P | ||
Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D) | |
Data Management | Software | Windows 7/8/10, 32or 64bit |
LIS System | Bi-direction,support HL7 protocol | |
Interface | LAN port access | |
Printer | External,multiple reporting mode available | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 300W |
Temperature | 10~30℃ | |
Water Consumption | Deionized water:12L/H | |
Humidity | 30-80% | |
Size & Weight | External Size (W*D*H) | 1167*780*1016mm |
Net Weight | 193kg | |
Package Size (W*D*H) | 1348*928*1399mm | |
Gross Weight | 267kg |
1. Reaction Tray
37±0.1℃, real-time monitor.
2. Mixer Probe
Teflon coating to avoid cross contamination.
3. Reagent Tray
2-8℃ cooling for 24 hours.
4. Sample Needle
Liquid level sensor function. Anti-collision function. Reagent volume real-time detection.
5. Washing Probe
Independent 7-step washing system.
6. Software
User-friendly software.
Product detail pictures:





Related Product Guide:
"Quality first, Honesty as base, Sincere service and mutual profit" is our idea, in order to develop continuously and pursue the excellence for OEM/ODM Manufacturer Fluorescence Measurement Microplate Reader - 400T / H Auto Chemistry Analyzer BK-400 – OLABO, The product will supply to all over the world, such as: Bolivia, Houston, Malta, Based on experienced engineers, all orders for drawing-based or sample-based processing are welcomed. We have won a good reputation for outstanding customer service among our overseas customers. We will continue to try the best to offer you good quality products and the best service. We are looking forward to serving you.