Parameter
Volume Range |
Increment |
Test Volume(ul) |
Accuracy error |
Precision error |
||
|
|
|
% |
μl |
% |
μl |
0.1-2.5μl |
0.05μl |
2.5 |
2.50% |
0.0625 |
2.00% |
0.05 |
1.25 |
3.00% |
0.0375 |
3.00% |
0.0375 |
||
0.25 |
12.00% |
0.03 |
6.00% |
0.015 |
||
0.5-10μl |
0.1μl |
10 |
1.00% |
0.1 |
0.80% |
0.08 |
5 |
1.50% |
0.075 |
1.50% |
0.075 |
||
1 |
2.50% |
0.025 |
1.50% |
0.015 |
||
2-20μl |
0.5μl |
20 |
0.90% |
0.18 |
0.40% |
0.08 |
10 |
1.20% |
0.12 |
1.00% |
0.1 |
||
2 |
3.00% |
0.06 |
2.00% |
0.04 |
||
5-50μl |
0.5μl |
50 |
0.60% |
0.3 |
0.30% |
0.15 |
25 |
0.90% |
0.225 |
0.60% |
0.15 |
||
5 |
2.00% |
0.1 |
2.00% |
0.1 |
||
10-100μl |
1μl |
100 |
0.80% |
0.8 |
0.15% |
0.15 |
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
10 |
3.00% |
0.3 |
1.50% |
0.15 |
||
20-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
20 |
3.00% |
0.6 |
1.00% |
0.2 |
||
50-200μl |
1μl |
200 |
0.60% |
1.2 |
0.15% |
0.3 |
100 |
0.80% |
0.8 |
0.30% |
0.3 |
||
50 |
1.00% |
0.5 |
0.40% |
0.2 |
||
100-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
100 |
2.00% |
2 |
0.70% |
0.7 |
||
200-1000μl |
5μl |
1000 |
0.60% |
6 |
0.20% |
2 |
500 |
0.70% |
3.5 |
0.25% |
1.25 |
||
200 |
0.90% |
1.8 |
0.30% |
0.6 |
||
1000-5000μl |
50μl |
5000 |
0.50% |
25 |
0.15% |
7.5 |
2500 |
0.60% |
15 |
0.30% |
7.5 |
||
1000 |
0.70% |
7 |
0.30% |
3 |
||
2-10ml |
0.1ml |
10ml |
0.60% |
60 |
0.20% |
20 |
5ml |
1.20% |
60 |
0.30% |
15 |
||
2ml |
3.00% |
60 |
0.60% |
12 |
Advantage
- Lightweight, ergonomic, low force design
- Digital display clearly reads volume setting
- The pipettes cover volume range of 0.1μl to 10ml
- Easy to calibrate and maintain with tool supplied
- Design helps avoid repetitive strain injuries
- Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate
- The low part is available for autoclaving
Product detail pictures:




Related Product Guide:
We can always satisfy our respected customers with our good quality, good price and good service due to we are more professional and more hard-working and do it in cost-effective way for New Fashion Design for Orbital Shaker Incubator Price - OLABO Single channel adjustable volume Mechanical pipette -toppette – OLABO, The product will supply to all over the world, such as: Manila, UK, Latvia, Besides strong technical strength, we also introduce advanced equipment for inspection and conduct strict management. All the staff of our company welcome friends both at home and abroad to come for visits and business on the basis of equality and mutual benefit. If you are interested in any of our items, please feel free to contact us for quotation and product details.