Parameter
Model | BK-400 | |
Overall | Throughput | 400 Test/hour |
Performance | Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry |
Certification | CE, ISO9001, ISO14001, ISO13485 | |
Sample Positions | 60 or 90 sample positions, built-in barcode system available | |
Sample & Reagent System | Reagent Positions | 45 refrigerated reagent positions |
Sample Volume | 2~70μl | |
Reagent Volume | 20~350μl | |
Probe | Teflon coating, with anti-collision function, liquid leveldetection function | |
Probe Washing | Automatic washing interior and exterior | |
Reagent Cooling | Refrigerated tray with independent switch | |
Temperature Control | 37±0.1℃, real-time monitoring | |
Reaction System | Cuvettes | 120reusable cuvettes, optical length 6mm |
Mixer Probe | Independent stirring | |
Washing | Automatic cuvettes washing | |
STAT function | YES | |
Optical System | Light Source | 12V/30W halogen-tungsten lamp |
Spectrophotometry | Post-spectral spectrophotometry | |
Wavelength | 340, 405, 450,510, 546, 578, 630,700nm | |
Absorbance | 0~3.0Abs | |
Calibration & QC | Calibration | Linear:K factor, 1-point, 2-point and multipoint linear |
Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P | ||
Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D) | |
Data Management | Software | Windows 7/8/10, 32or 64bit |
LIS System | Bi-direction,support HL7 protocol | |
Interface | LAN port access | |
Printer | External,multiple reporting mode available | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 300W |
Temperature | 10~30℃ | |
Water Consumption | Deionized water:12L/H | |
Humidity | 30-80% | |
Size & Weight | External Size (W*D*H) | 1167*780*1016mm |
Net Weight | 193kg | |
Package Size (W*D*H) | 1348*928*1399mm | |
Gross Weight | 267kg |
1. Reaction Tray
37±0.1℃, real-time monitor.
2. Mixer Probe
Teflon coating to avoid cross contamination.
3. Reagent Tray
2-8℃ cooling for 24 hours.
4. Sample Needle
Liquid level sensor function. Anti-collision function. Reagent volume real-time detection.
5. Washing Probe
Independent 7-step washing system.
6. Software
User-friendly software.
Product detail pictures:





Related Product Guide:
We have our own sales team, design team, technical team, QC team and package team. We have strict quality control procedures for each process. Also, all of our workers are experienced in printing field for Factory Price For Urine Analysis Biochemistry - 400T / H Auto Chemistry Analyzer BK-400 – OLABO, The product will supply to all over the world, such as: Mauritius, Comoros, Miami, Based on our automatic production line, steady material purchase channel and quick subcontract systems have been built in mainland China to meet customer's wider and higher requirement in recent years. We have been looking forward to cooperating with more clients worldwide for common development and mutual benefit!Your trust and approval are the best reward for our efforts. Keeping honest, innovative and efficient, we sincerely expect that we can be business partners to create our brilliant future!