Excellent quality Elisa Wash Bottle - 1200T / H Auto Chemistry Analyzer BK-1200 – OLABO
Excellent quality Elisa Wash Bottle - 1200T / H Auto Chemistry Analyzer BK-1200 – OLABO Detail:
Parameter
Model | BK-1200 | |
Overall Performance | Throughput | 800 Tests/hour, 1200 Tests/hour with ISE(Optional) |
Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry | |
Certification | CE, ISO9001, ISO14001, ISO13485 | |
Sample & Reagent System | Sample Positions | 150 sample positions |
Reagent Positions | 180 reagent positions | |
Sample Volume | 2~70μl | |
Reagent Volume | 20~350μl | |
Probe | Teflon coating ,with anti-collision function, liquid leveldetection function | |
Probe Washing | Automatic washing interior and exterior | |
Reagent Cooling | Refrigerated tray with independent switch | |
Reaction System | Temperature Control | Water bath, 37±0.1℃, real-time monitoring |
Cuvettes | 160 reusable cuvettes, optical length 6mm | |
Mixer Probe | Independent stirring | |
Washing | Automatic cuvettes washing | |
STAT function | YES | |
Optical System | Light Source | 12V/20W halogen lamp/fiber bragg grating optional |
Spectrophotometry | Post-spectral spectrophotometry | |
Wavelength | 12 wavelengths from 340~800nm | |
Absorbance | -0.3~33.0Abs | |
Calibration & QC | Calibration | Linear:K factor, 1-point ,2-point and multipoint linear |
Non-Linear:Spline,Polygon, Index, Ogarithm, Logit-4P, Logit-5P | ||
Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D) | |
Data Management | Software | Windows 7/8/10 , 32or 64bit |
LIS System | Bi-direction,support HL7 protocol | |
Interface | LAN port access | |
Printer | External,multiple reporting mode available | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 1500W |
Temperature | 10~30℃ | |
Water Consumption | Deionized water:20L/H | |
Humidity | 30-80% | |
Size & Weight | External Size (W*D*H) | 1360*791*1157mm |
Net Weight | 286kg | |
Package Size (W*D*H) | 1718*929*1434mm | |
Gross Weight | 390kg |
Advantages
1. Reaction Tray
37±0.1℃, water bath.
2. Mixer Probe
Dual mixer, auto frequency adjustment.
3. Reagent Tray
2-8℃ cooling for 24 hours.
4. Sample Probe
Liquid level sensor function. Anti-collision function. Reagent volume real-time detection.
5. Washing Probe
Dual design, high efficiency.
6. Software
User-friendly software.
Product detail pictures:






Related Product Guide:
Adhering to your principle of "quality, assistance, performance and growth", we have now gained trusts and praises from domestic and international customer for Excellent quality Elisa Wash Bottle - 1200T / H Auto Chemistry Analyzer BK-1200 – OLABO, The product will supply to all over the world, such as: Nepal, Japan, Germany, Immediate and specialist after-sale service supplied by our consultant group has happy our buyers. Detailed Info and parameters from the merchandise will probably be sent to you for any thorough acknowledge. Free samples may be delivered and company check out to our corporation. n Morocco for negotiation is constantly welcome. Hope to get inquiries type you and construct a long-term co-operation partnership.

This company can be well to meet our needs on product quantity and delivery time, so we always choose them when we have procurement requirements.