8 Year Exporter Clinical Chemistry Analyzer - 400T / H Auto Chemistry Analyzer BK-400 – OLABO
8 Year Exporter Clinical Chemistry Analyzer - 400T / H Auto Chemistry Analyzer BK-400 – OLABO Detail:
Parameter
Model | BK-400 | |
Overall | Throughput | 400 Test/hour |
Performance | Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry |
Certification | CE, ISO9001, ISO14001, ISO13485 | |
Sample Positions | 60 or 90 sample positions, built-in barcode system available | |
Sample & Reagent System | Reagent Positions | 45 refrigerated reagent positions |
Sample Volume | 2~70μl | |
Reagent Volume | 20~350μl | |
Probe | Teflon coating, with anti-collision function, liquid leveldetection function | |
Probe Washing | Automatic washing interior and exterior | |
Reagent Cooling | Refrigerated tray with independent switch | |
Temperature Control | 37±0.1℃, real-time monitoring | |
Reaction System | Cuvettes | 120reusable cuvettes, optical length 6mm |
Mixer Probe | Independent stirring | |
Washing | Automatic cuvettes washing | |
STAT function | YES | |
Optical System | Light Source | 12V/30W halogen-tungsten lamp |
Spectrophotometry | Post-spectral spectrophotometry | |
Wavelength | 340, 405, 450,510, 546, 578, 630,700nm | |
Absorbance | 0~3.0Abs | |
Calibration & QC | Calibration | Linear:K factor, 1-point, 2-point and multipoint linear |
Non-Linear: Spline, Polygon, Index, Ogarithm, Logit-4P, Logit-5P | ||
Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D) | |
Data Management | Software | Windows 7/8/10, 32or 64bit |
LIS System | Bi-direction,support HL7 protocol | |
Interface | LAN port access | |
Printer | External,multiple reporting mode available | |
Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 300W |
Temperature | 10~30℃ | |
Water Consumption | Deionized water:12L/H | |
Humidity | 30-80% | |
Size & Weight | External Size (W*D*H) | 1167*780*1016mm |
Net Weight | 193kg | |
Package Size (W*D*H) | 1348*928*1399mm | |
Gross Weight | 267kg |
1. Reaction Tray
37±0.1℃, real-time monitor.
2. Mixer Probe
Teflon coating to avoid cross contamination.
3. Reagent Tray
2-8℃ cooling for 24 hours.
4. Sample Needle
Liquid level sensor function. Anti-collision function. Reagent volume real-time detection.
5. Washing Probe
Independent 7-step washing system.
6. Software
User-friendly software.
Product detail pictures:





Related Product Guide:
We aim to find out high quality disfigurement in the generation and provide the most effective services to domestic and abroad clients wholeheartedly for 8 Year Exporter Clinical Chemistry Analyzer - 400T / H Auto Chemistry Analyzer BK-400 – OLABO, The product will supply to all over the world, such as: Malta, Malta, Seattle, We have won a good reputation among overseas and domestic clients. Adhering to the management tenet of "credit oriented, customer first, high efficiency and mature services", we warmly welcome friends from all walks of life to cooperate with us.