1200T / H Auto Chemistry Analyzer BK-1200
Parameter
| Model | BK-1200 | |
| Overall Performance | Throughput | 800 Tests/hour, 1200 Tests/hour with ISE(Optional) |
| Analysis Method | End-point, Fixed-time, Rate(Kinetic), Turbidimetry | |
| Certification | CE, ISO9001, ISO14001, ISO13485 | |
| Sample & Reagent System | Sample Positions | 150 sample positions |
| Reagent Positions | 180 reagent positions | |
| Sample Volume | 2~70μl | |
| Reagent Volume | 20~350μl | |
| Probe | Teflon coating ,with anti-collision function, liquid leveldetection function | |
| Probe Washing | Automatic washing interior and exterior | |
| Reagent Cooling | Refrigerated tray with independent switch | |
| Reaction System | Temperature Control | Water bath, 37±0.1℃, real-time monitoring |
| Cuvettes | 160 reusable cuvettes, optical length 6mm | |
| Mixer Probe | Independent stirring | |
| Washing | Automatic cuvettes washing | |
| STAT function | YES | |
| Optical System | Light Source | 12V/20W halogen lamp/fiber bragg grating optional |
| Spectrophotometry | Post-spectral spectrophotometry | |
| Wavelength | 12 wavelengths from 340~800nm | |
| Absorbance | -0.3~33.0Abs | |
| Calibration & QC | Calibration | Linear:K factor, 1-point ,2-point and multipoint linear |
| Non-Linear:Spline,Polygon, Index, Ogarithm, Logit-4P, Logit-5P | ||
| Quality Control | Real-time QC, Westgard multi rule, Cumulative Sum Check,Twin Plot(2D) | |
| Data Management | Software | Windows 7/8/10 , 32or 64bit |
| LIS System | Bi-direction,support HL7 protocol | |
| Interface | LAN port access | |
| Printer | External,multiple reporting mode available | |
| Working Conditions | Power Supply | AC220V±10%,50/60Hz, 110V±10%,60Hz, 1500W |
| Temperature | 10~30℃ | |
| Water Consumption | Deionized water:20L/H | |
| Humidity | 30-80% | |
| Size & Weight | External Size (W*D*H) | 1360*791*1157mm |
| Net Weight | 286kg | |
| Package Size (W*D*H) | 1718*929*1434mm | |
| Gross Weight | 390kg | |
Advantages
1. Reaction Tray
37±0.1℃, water bath.
2. Mixer Probe
Dual mixer, auto frequency adjustment.
3. Reagent Tray
2-8℃ cooling for 24 hours.
4. Sample Probe
Liquid level sensor function. Anti-collision function. Reagent volume real-time detection.
5. Washing Probe
Dual design, high efficiency.
6. Software
User-friendly software.